6 07/05• Click on the Laser button to open the Laser Control window.• Select the appropriate Laser Unit by clicking onthe name of it.• Click on the Standby button to switch requiredlaser(s) to Standby.• When status is Ready click on On button.• Set Output [%] so that the Tube Current isabout 4 A (50% of the output power).Setting the microscopeChanging between direct observation or laser scanningThe VIS, TV and LSM buttons switch the beam path and indicate which beam path has been set in thebinocular tube of the microscope:• Click on the VIS button to set the microscope for direct observation via the eyepiecesof the binocular tube, lasers are off.• Click on the TV button to set the microscope camera observation (if connected) viacamera adapter of the binocular tube.• Click on the LSM button to set the microscope screen observation via laser excitationusing the LSM 510 and software evaluation.Setting the microscope and storing the settings• Click on the VIS button for direct observation.• Click on the Micro button in the Acquire subordinate toolbar to open theMicroscope Control window of the used microscope.The Microscope Control window appears (Fig. 10).Fig. 9 Laser Control window