page 35 of 64ConfigurationIn the ”Configuration” menu the device parameters are preselected.Preselections+ Use the selection knob to select the parameters.➨ Switch the DCP 3000 on.➨ Press the selection knob.➨ Turn the selection knob to choose a menu and confirm by pressing theselection knob.’’Configuration’’Settings for• Adjustment (of the pressure transducer)• RS-232 (serial interface)• Sensors (configuration and switching between several pressure trans-ducers)• Display (brightness and contrast of the display, language, sound,.....)• Setpoint (valves, digital I/O interface)• Log Data (Off, 1 . . . 3600 seconds)• Graphic+ Adjustment: Adjustment of the pressure transducer under vacuumand/or at atmospheric pressure. See also sections, “Readjustment ofVSK 3000”, pg. 54. / “Readjustment of VSP 3000”, pg. 56.Adjustment is carried out at two adjustment points: at atmosphericpressure and under vacuum.If no adjustment is possible in the actual pressure range, ”---- Torr” isdisplayed.+ RS-232: Configuration of the interface, setting of parameters and com-mands, see section “Interface parameters”, pg. 43.Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”,”7-O-1” or ”7-E-1”, Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” andremote on ”On” or ”Off”.Automatic sending of pressure values ”off” (only on demand) or in atime interval of 1 .... 3600 seconds.