page 46 of 64Command Operation Response DescriptionIN_PV_1 current pressureXXXX mbar/Torr/hPa*X.XXE±X mbar/Torr/hPa*unit according to preselectionsIN_PV_Sxcurrent pressureof pressure trans-ducer xXXXX mbar/hPa/Torr*X.XXE±X mbar/Torr/hPa*pressure of pressure transducer x(order of numbering according to dis-play in ”Sensors” menu)IN_PV_Xcurrent pressurevalues of all con-nected pressuretransducersXXXX.X XXXX.X...mbar/Torr/hPa*X.XXE±X X.XXE±X...mbar/Torr/hPa*pressure values of all connected pres-sure transducersIN_PV_D data readout XXX.XX:XX:XX X...mbar/Torr/hPa* read out data from DCP 3000 memoryIN_CFG device setpreselectionsyXXXXXXX0XXXXXX1XXXXXX2XXXXXXX0XXXXXX1XXXXXXX0XXXXXX1XXXXXXX0XXXXXX1XXXXXXXyXXXXXXXyy: 0......D: language** (hexadecimal)pressure unit mbarpressure unit Torrpressure unit hPaacoustic signal offacoustic signal onventing valve not connectedventing valve connectedfault indicator not connectedfault indicator connectedy: 1......8: number*** of active pressuretransducery: 1......8: pressure transducer quantityIN_ERR fault status0XXXX1XXXXX0XXXX1XXXXX0XXXX1XXXXX0XXXX1XXXXX0XXXX1no fault at venting valvefault at venting valveno overpressureoverpressureno fault at pressure transducerfault at pressure transducerno external faultexternal faultlast interface command correctlast interface command incorrectIN_VER version DCP 3000 Vx.xx software versionIN_SP_1 time interval forsending XX:XX mm:ss (minutes:seconds);time interval for sendingIN_SP_2 time interval forrecording XX:XX mm:ss: (minutes:seconds);time interval for recordingRead commands