page 39 of 64determined and indicated between two VSK 3000 pressure transducersconnected to the DCP 3000.+ This requires one of the two pressure transducers to be defined asreference (e.g., VACUU•BUS address ”Ref. 1”).+ In addition, both pressure transducers have to be configured to cor-responding VACUU•BUS addresses (VSK x and Ref. x). If the two ad-dresses do not correspond (VSK x and Ref. y), no differential pressurewill be determined or displayed. Then it is necessary to reconfigurethe address of one pressure transducer to enable differential pressuremeasurement (see section ”Reconfiguring the pressure transducer”below).➨ The displayed pressure value is: (”pressure value of Ref. x” minus”pressure value of VSK x”) TorrReconfiguring the pressure transducer VSP 3000 or VSK3000Configuration of the MPT 100: See instructions for use of MPT 100.Reconfiguring the VACUU•BUS address of a pressure transducer:1. Switch the vacuum gauge DCP 3000 off.2. Preferably connect only the pressure transducer to be reconfigured tothe DCP 3000.3. Keep the ”VENT” key pressed and switch the DCP 3000 on (”ON/OFF” key). Release the ”VENT” key.4. Select the menu item ”Vacuubus” by turning and pressing the selec-tion knob.5. The first component detected by the VACUU•BUS is displayed (e.g.,VSK 1). If necessary, select the pressure transducer to be reconfig-ured by turning and pressing the selection knob.6. Press the ”VENT” key and keep pressed, then press additionally theselection knob. A frame will replace the inverse display of the compo-nent. Release the ”VENT” key.7. Turn the selection knob until the address, to which you want to recon-figure the pressure transducer (e.g., VSK 2 or Ref.1), is displayed.8. Press the selection knob. If the configuration was successful, the menuis left and the DCP 3000 will restart.