Operation ECA 45024 Instruction 0024-94008. With the analyzer set up per Section 2.4 Performing a Combustion Test,press ENTER key to begin testing and logging.«LOG«««(0025)««NGAS«This display shows that the analyzer is logging and currently saving datato internal memory location 25. Observe that the memory locationnumber will increment as data is being saved. Note that if all memorylocations are used before the end of the logging duration, then “(xxxx)”appears in place of the memory location number, indicating a memoryover-range condition.«LOG«««(-PC-)««NGAS«This display shows that the analyzer is logging data, and that the data isbeing sent to only a personal computer.Tip: Logging can be stopped without loosing data that has already beencollected by pressing the ESC key.9. At the conclusion of the logging process, the message “Logging Complete”will appear. And if data was being stored in the analyzer’s internalmemory, then the message “Data in memory nnnn to nnnn” will alsoappear — where nnnn denotes the internal memory locations thatcontain the logged data. Press ENTER to leave logging and display theCombustion Test HOLD screen.2.12.2 Turning Data Logging OffData logging is automatically turned off . . .• at the end of the logging process when the ENTER key pressed.• if the logging process is interrupted by pressing the ESC key.Note: If the logging process is interrupted, it cannot be resumed without firstturning logging back on per Section 2.12.1.