Müse Parameter Descriptions319FHR Signal LossThe FHR Signal Loss parameter controls the overall quality of the fetal heart rate signal to emulatequality of the probe installation. This parameter indicates the relative amount of signal loss as apercentage. 0% is a clean signal with no loss. 100% results in no signal. Intermediary values createprogressively larger gaps of data on the graph. The FHR widget displays “-“ when there is no valid data.This parameter is applied, regardless of the probe type. Instructors can create various FHR signal qualityscenarios by using this control and the probe type selection.Default: 0Range: 0 - 100UA Signal LossThe UA Signal Loss parameter controls the overall quality of the uterine activity signal. When set to“yes”, the uterine activity signal is set to 0 regardless of the probe type.Default: NoTOCO OffsetThe TOCO Offset parameter is applicable only if the UA probe selected is TOCO. This parameter is usedto shift the uterine activity signal by adding/removing a fixed pressure offset to the uterine pressure. Itis useful to emulate conditions such as a tight UA probe belt. The effect is only from the change,previous historical data is not changed.Default: 0Range: -100 - 100MHR GraphThe MHR Graph parameter activates the display of the maternal heart rate over the FHR display, whenset to “on”. The maternal heart rate graph appears as a blue line.Default: OffMHR Overrides FHRThe MHR Overrides FHR parameter activates the condition whereby the FHR probe picks-up thematernal heart beat rate, when set to “on”.Default: OffMSPO2 Printing FrequencyThe MSPO 2 Printing Frequency parameter activates printing of the maternal SpO 2 value on the stripat the specified frequency selected. Printing occurs only if there is a connected SpO 2 probe.Frequency selections include, in minutes: 1, 2, 3, 5, 10, 15, 30, 60, 90Default: 10 minutes