ES601Plus OPERATING MANUALFig u re 41 - Example of multiple PE test result.5.14 Viewing Test Data and Report Output5.14.1 Viewing a RecordPreviously-saved Test Records may be recalled from the flash memory as described in theimmediately preceding section. The current contents of the Clip Record may also be viewed whiletesting is in progress, via the VIEW RECORD button on the MANUAL MODE menu. In eithercase, test data is shown on the LCD as a multi-page report which is referred to as a "Test Record".The term "page" in this case refers not to a printed page, but to the amount of data that can beshown on the display of ES601Plus at one time; this is typically 20 lines or less.If no data has been obtained when VIEW RECORD is pressed, or if the Clip Record has beenerased by changing the safety standard or saving a Test Record, ES601Plus will display “THERE ISNO MEASUREMENT DATA TO VIEW”, and will automatically return to the MANUALMODE menu.If a safety test was performed, page 1 of the Test Record will include results for line voltage, loadcurrent and insulation resistance. Page 2 of the Record, an example of which appears below, will listresults of the protective earth (PE) test if this test was performed. For multiple PE tests, this pagemay list up to ten resistance measurements according to the test protocol conveyed by theinformation appearing near the top right of the page. For a single PE test, only one resistance willappear in the "VALUE" column. If no point-to-point measurements were performed, then the"OFFSET:" will be represented as "-.---".Manual Testing/Chapter 5 # Page 85