ES601Plus OPERATING MANUALFig u re 71 - Test Procedure selection menu.6.1.4 Test ProceduresA Test Procedure can integrate visual inspection, safety testing, ECG testing, and performancetesting activities into one comprehensive procedure targeted at a specific kind of medical device, forexample, a patient monitor, defibrillator, anaesthesia gas machine, infusion pump, etc. The flashmemory of ES601Plus provides storage of up to 500 Test Procedures.The Test Procedure Editor is supplied as an optional part of the ES601pc software package.Provided ES601Plus has the Test Procedure accessory option installed, the Test Procedure Editorof ES601pc will allow the user to program ES601Plus with Test Procedures. The Test ProcedureEditor of ES601pc is described in Chapter 8. ES601Plus menus associated with selecting andrunning Test Procedures created by the Test Procedure Editor are discussed below.If the optional accessory feature associated with Test Procedures has not been installed inES601Plus, this will be indicated on the main menu for AUTOSEQUENCE MODE as "(NOTINSTALLED)". In this case, the Test Procedure menu will not be selectable. Otherwise, selectingTEST PROCEDURES from the AUTOSEQUENCE MODE menu opens the Test Proceduresmenu.Automated Testing/Chapter 6 # Page 114