SEFELEC1500M Instructions 22 V1.065. COMMISSIONINGWARNINGRisk of electric shock!• The equipment must be commissioned, used and maintained by qualifiedpersonnel.• All precautions for using equipment connected to mains supplies must betaken when it is used.• In particular, the equipment must be earthed through the power cord andthrough the dedicated earth connection on the rear panel (see16.1 sectionZ2 note)• Always ensure that the high voltage presence light is off before connectingor disconnecting a wired assembly to be tested.5.1. CONTENT WHEN DELIVEREDWhen the measuring device is delivered, the package should contain the following items:• SEFELEC1500M megohmmeter device• 1 x 2P+E 16 A power cord, 1.5 m in length.• 4 x 10-points plugs of the screw terminal type with green cover• 1 x printed Complete user manual for the device.• 1 x printed guide for safety rules in high voltage test• 1 x CE mark conformity certificate (included at the end of this manual)And according to your order:• The SEFA-SE15, SEFA-HV15, …measuring accessories• WinPass-MX software.• Measurement data documentPlease contact EATON if any items are missing. See the chapter After-sales service.5.2. SETTING UP THE VARIOUS CONNECTORS AND CORDSBefore turning the measuring device on, check that the following connectors and accessoriesare correctly connected and locked in position:• The 10-pin C5 safety circuit plug should be connected to the 10-pin safety circuit socket (notethat these sockets are keyed)• Power cord on the rear panel and on a 16A mains socket with earth connection.