Determine nominal weight Display1. Place nominal weight (e.g. 100 g) on the pan.2. Press TARE-key: the display indicates zero.3. Remove nominal weight. It is displayed with a negative sign.100,0 g0,0 g- 100,0 gDetermine deviations Display1. Place sample to be compared with the nominal weight onthe pan.2. If the object is heavier (e.g. 104,5 g) the deviation isdisplayed with a positive sign.3. If the object is lighter (e.g. 97,5 g) the deviation is displayedwith a negative sign.4,5 g-2,5 g4.6 Piece countingIn order to select the parts counting mode the following keys have to be pressed:Selection of the reference piece quantity Display1. Press MODE-key, "CAL? " appears.2. Press MODE-key, "PRINTER?" appears.3. Press MODE-key, "Count?" appears.4. Confirm with YES-key5. "REF=1" appears.The desired reference can be selected by the MODE-key.6. The following references can be selected: 1,5,10 and 20pieces.=> The desired reference to be confirmed by theYES-key.7. Return with the TARE-key to the weighing mode.CAL?PRINTER?Count?Count?xREF=1?0,0 gCounting and determine reference (e.g. reference = 10 pcs.) Display1. Press TARE-key to set the display to zero.2. Place 10 pieces on the weighing plate.3. Press REF-key. The balance is ready to count.The counting accuracy can be improved by the followingproceedings:4. Double the weighing object: Add (about) another 10 pieces.The balance calculates the new reference and displays thenew quantity, wait for shortly flashing "opt".5. Double again (see point 4)6. The balance is ready to count.0,0 g(Weight)10 PCSopt 20 PCSopt 40 PCS30 572/KB/DS-BA-defsi-0246