3.4 Description of display promptsPrompt Meaning==== OVERLOAD====or===================Overload:Weighing range exceededThe display shows minus values=================== Underload<< Counting/percent mode: the parts weigh too light.Opt Counting mode: The balance calculates automatically a betterreference.X Confirmation sign for selected option in the MODE-program4 Operation4.1 Turn the balance on and offTurn the balance on and off Display1. Press ON/OFF-key2. The display indicates3. The weighing range is indicated, e.g.:4. The display indicates5. and finallyAll segments- - -6000,0 g- - -- - - P 7.40 - - -- - - - - - - - - - - -0,0 g4.2 Adjusting (CAL)4.2.1 Adjustment of non verifiable instrumentsPrior to first use, it is necessary to adjust the balance at the place intended for use.The accuracy of the balance can be checked and adjusted at any time with an adjustingweight. Take care of stable environmental conditions, see "Important Information". Awarming up time of 15 minutes is necessary.Adjusting (CAL) Display1. Press TARE-key to put the display to zero:2. Press MODE-key, "CAL" appears:3. Press YES-key t.4. Wait until "PLACE CAL.-WEIGHT" appears:5. Place the adjusting weight on the weighing plate.0,0 gCAL?CALIBRATION ZEROPLACE CAL.-WEIGHTCALIBRATION SPANAvoid any vibration of the balance!6. Adjusting is finished with the steady display of the adjustingweight, e.g. 500,0 gKeep the adjusting weight near the balance for daily control of the balance’s accuracy.572/KB/DS-BA-defsi-0246 27