AES/AEJ-BA-e-0912 25• When the PRINT-key is pressed during the adjusting procedure, [stop] willbe displayed and adjustment interrupted. The balance returns to weighingmode.• You can cancel the adjusting procedure by pressing any key.• The following error messages may be displayed during adjust-ment/adjustment test.1-Err Incorrect adjusting weight2-Err Divergence last external adjustment > 1%3-Err During the adjustment there was one weight on the weighing plate4-Err Divergence from last internal adjustment > 1%A-Err Internal adjustment automatics defectiveAdjustment protocols can be issued by connecting an optionalprinter, see chpt. 15 "printer operation".[ ] is displayed during printing.9.5 Advice CAL/Auto-CAL (KERN PEJ only)Automatic adjustment is started or required every 4 hours, or when a temperaturechange of 5 ºC occurs, depending on the setting in the menu, see chpt. 10.1.1.Advice CAL = menu setting [ E. ADC. 1 ]When “Advice CAL“ is set, the required adjustment will merely be requested andmanual adjustment must be carried out.“CAL” will start flashing if adjustment is required (after 4 h/ ± 5ºC). “CAL" will keepflashing until the adjustment is started.For AEJ models this is carried out with an internal adjustment weight (chpt. 9.1), forAES models with an external adjustment weight (chpt. 9.3).Advice CAL = menu setting [ E. ADC. 2 ]Automatic adjustment is carried out when “Auto-CAL“ is set and an adjustment due(after 4 h / ± 5ºC).“CAL” will start flashing 10 minutes before the start of automatic adjustment withinternal adjusting weight. Complete current weighing in processes within this timeframe. Automatic adjustment will not be started until the weighing balance isunloaded and a successful stability display has appeared.It is only possible to enable the Auto-Cal function in menu setting[ x CA. 1 ], see chpt. 10.1.2.