OperationITS-BA-e-072014User manual ME-Nr.:22017077 01/072.5 Calling up the gross weight and tare value1. Press .The gross weight is displayed for 2 seconds.2. Press again while the gross weight is displayed.The tare value is displayed. After a few seconds, the scale changes back to the netweight display.2.6 Displaying weight values with a higher resolution➜ Press .The current weight value is displayed for 2 seconds in a higher resolution. Thescale then changes back to the normal resolution.2.7 Displaying the capacity availableThe scale has a graphic display of the scale capacity available. The bar indicates howmany per cent of the scale capacity is already occupied and what capacity is stillavailable. In the example, approx. 65 % of the scale capacity is occupied.2.8 Dynamic weighingWith the dynamic weighing function, it is possible to weigh restless weighing samplessuch as live animals. If this function is activated, the symbol appears in the dis-play.With dynamic weighing, the scale calculates the mean value from 56 weighing oper-ations within 4 seconds.With manual start PrerequisiteAVErAGE -> MAnuAL is selected in the menu.The weighing sample must be heavier than 5 scale divisions.1. Place the weighing sample on the scale and wait until it has stabilized.2. Press to start dynamic weighing.During dynamic weighing, horizontal segments appear in the display, and thedynamic result is then displayed with the symbol *.3. Unload the scale to be able to start a new dynamic weighing operation.