Event and error messagesITS-BA-e-072042User manual ME-Nr.:22017077 01/076 Event and error messagesError Cause RemedyDisplay Dark • Back lighting set too dark ➜ Set back lighting (b.LIGHt)brighter• No mains voltage ➜ Check mains• Unit switched off ➜ Switch on unit• Mains cable not plugged in ➜ Plug in mains plug• Brief fault ➜ Switch device off and back onagainInsufficient load • Load plate not on the scale ➜ Place load plate on the scale➜ Set to zero• Weighing range not reachedOverload • Weighing range exceeded ➜ Unload scale➜ Reduce preload• Result not yet stable ➜ If necessary adjust vibrationadapter or weigh dynamically• Function not permissible ➜ Unload scale and set to zero• Zeroing not possible with over-load or insufficient load➜ Unload scale• Average piece weight too low ➜ Select and place larger number ofreference parts on the scale• No valid value from the referencescale➜ Check cable connection betweenthe units➜ Check interface settings• No calibration ➜ Unplug the mains plug then plugit back in; switch unit off and thenback on in battery mode➜ Calibrate scale➜ Contact your dealer or local rep-resentative• Average piece weight too low ➜ Counting is not possible on thisscale with this average pieceweight