ITS-BA-e-072037Interface descriptionUser manual ME-Nr.:22017077 01/075 Interface description5.1 SICS interface commandsThe terminal supports the command set MT-SICS (METTLER TOLEDO Standard Inter-face Command Set). With SICS commands, it is possible to configure, query andoperate the terminal from a PC. SICS commands are divided up into various levels.5.1.1 Available SICS commandsCommand MeaningLEVEL 0 @ Reset the scaleI0 Inquiry of all available SICS commandsI1 Inquiry of SICS level and SICS versionsI2 Inquiry of scale dataI3 Inquiry of scale software versionI4 Inquiry of serial numberS Send stable weight valueSI Send weight value immediatelySIR Send weight value repeatedlyZ Zero the scaleZI Zero immediatelyLEVEL 1 D Write text into displayDW Weight displayK Keyboard checkSR Send and repeat stable weight valueT TareTA Tare valueTAC Clear tareTI Tare immediatelyLEVEL 2 C2 Calibrate with the external calibration weightC3 Calibrate with the internal calibration weightI10 Inquire or set scale IDI11 Inquiry of scale typeP100 Print out on the printerP101 Print out stable weight valueP102 Print out current weight value immediatelyPWR Power On/Off