Instruction ManualOCM FOCM F - Rev. 04 as of 18.02.2014 Page 479.4 Parameter Menu (PAR)This menu is the most important regarding the unit settings. It nevertheless issufficient in most cases to set only some essential parameters, which usuallyare:− Channel shape− Channel dimensions− Sensor types− Analog output (function, measurement range and measurement span)− Relay output (function and values)All other functions are additions which are required in special cases only.(Regulator mode or for special hydraulic applications). These settings are nor-mally made with the help of our service personnel or by an authorised ex-pertcompany.The parameter menu >PAR< includes eight partially very extensive submenuswhich are described individually on the following pages.9.4.1 Parameter Menu „Measurement Place“This menu is one of the most important basic menus for parameter setting as themeasurement place is going to be defined here.The menu cannot be indicated completely due to restricted display space. Simi-lar to many well-known PC applications, this is readily identifiable from the blackbars on the right-hand side of the screen.Name NIVUS recommends to coordinate and to define names according to namesstated in the respective documents. Names may contain up to 15 letters.After the submenu >Name< has been selected the basic setting „NIVUS“ willcome up. There is a cursor blinking below the first digit.Fig. 9-17 submenu – nameUse "Up“ and "Down“ keys to scroll the menu.Underneath the measurement place name you can find a table with 20 lines con-taining all uppercase and lowercase letters as well as a large number of specialcharacters (see Fig. 9-18). Use the arrow keys >up< and >down< to jump across2 lines up or down at each key action.