Instruction ManualOCM FOCM F - Rev. 04 as of 18.02.2014 Page 799.6.4 CAL Menu „Digital outputs“Selecting the >Digital outputs< menu requires to enter the PIN (2718) onceagain. This ensures to prevent unauthorised persons from performing simula-tions during operation.Choose the desired relays to be simulated by using the >up< and >down< arrowkeys. Use ENTER to either turn the chosen relay ON or OFF directly. The relayswill de-energise if you exit the menu.Fig. 9-80 Relay simulation9.6.5 CAL Menu „Simulation“After entering the PIN code >2718< select between level, velocity and mediumtemperature by using the >up< and >down< arrow keys. Pressing >right< or>left< will increase or decrease the simulated flow velocity or temperature valuein steps of 1 cm or 0.1 °C.Using ENTER enables to enter the desired simulation value directly.The flow value which has been calculated by means of the simulated readingswill be indicated on the bottom line of the screen. Relays which might have beenset will switch and programmed mA outputs supply according current valuessimultaneously. Use >up< or >down< keys to indicate output values in the bot-tom line of the display.1 Simulated height/level2 Simulated flow velocity3 Simulated medium temperature4 Calculated simulated flow value5 Programmed relay activated by simulation6 Analog output signalsFig. 9-81 Simulation mode