EN-8 RANGER®7000 SCALES2.8 Separating the Terminal from the Weighing Base1. To detach, press both the Release buttons inward (both at the same time) and gently pull the Terminaltowards you (outward) until the Terminal is detached. These Release buttons disengage the two hooksholding the Terminal to the Base. A cable is attached to the Terminal. Take care to not damage ordisconnect this cable.2. To reattach the Terminal, press in the two Release buttons and slide the Terminal into the Base until theTerminal hooks click and engage to hold the Terminal in place.Release ButtonsBase and Terminal2.9 Terminal MountingIf desired, the Terminal may be mounted to a wall or table using fasteners (not supplied) that are appropriate for thetype of mounting surface.2.10 Initial CalibrationWhen the Scale is first installed, and when it is moved to another location, it must be calibrated to ensure accurateweighing results.2.10.1 Internal calibrationR71MHD models have built in AutoCal which can calibrate the scale automatically and does not require calibrationmasses. If preferred, the scale can be manually calibrated with external masses. Have the appropriate calibrationmasses available before beginning calibration. Refer to the Calibration Section for masses and calibrationprocedure.2.10.2 External calibrationR71MD models can only be manually calibrated with external masses.50 mm /1.97 inMounting holes