● Error codeErrorcode Code type Description01 Command error The command is undefined.02 Address error The start address is larger than the end address or theaddress is larger than 999999 when the RDD or WDDcommand is executed.The address length has not reached the prescribedlength when the RDD or WDD command is executed.03 Data error The data length does not correspond to the command.The data length has not reached the prescribed length.04 BCC error BCC check was not conformable.11 CommunicationerrorA parity error occurred during data reception.A framing error occurred during data reception.An overrun error occurred during data reception.21 Control flow error The system is in setting mode.22 Execution error Calibration or analog scaling is not executable.31 Bufferingcondition error 1An attempt was made to make a buffering settingchange without stopping buffering.32 Bufferingcondition error 2 An attempt was made to start buffering with aninadequate buffering setting.33 Bufferingcondition error 3Data was read after buffering operation started.Data was read while the system was not in theaccumulation completed status.Data in excess of the final data point was specified andread.• If an abnormal response is received, check if there are any errors such astransmission data errors. Correct such errors, if any. If an error occursregardless of the absence of mistakes, it may be a temporal phenomenoncaused by noise or other external reasons or the external host device mayhave a system error. Turn the sensor head or system turned OFF and ON.Ramco National800-280-6933 | nsales@ramcoi.comwww.panasonicsensors.com