906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide A-1Appendix ASpecificationsGeneralParameter Operating RangeOperating Modes EMV+ ® and Eagle II™: AC,SIMV, CPAP with and without Pressure Support,BL Modes, and Leak Compensation for Active and Noninvasive ventilation.AEV®: AC, CPAP (with and without Pressure Support), BL Modes, and LeakCompensation for Noninvasive ventilation.)Breath Target Volume or PressureFlow Rate 0 to 100 LPM at 40 cm H 2OBreath Rate 1 to 80 BPM ±1 BPM over the intervalSetting Resolution: 1 BPMMeasurement: 1 to 90 BPM ± 1 BPM over the intervalInspiratory Time (Ti) Setting: 0 to 3 ± s for I:E ratio > 1:1.0, 0 to 5 ± s for E:I ratio <= 1.0:1Setting Resolution: 0.05 sTidal Volume Setting: 50 to 2000 ml ATPD ± (5 ml +10% setting)Setting Resolution: 10 mlMeasurement: 0 to 9999 ml ATPD ± (5 ml +10% setting)FIO 2 21 to 100% (± 3% of full scale ± 10% of setting)PEEP/EPAP Setting: 0 to 30 cm H 2O ± (2 cm H2O + 8% of reading)Setting Resolution: 10 mlMeasurement: 0 to 30 cm H 2 O ± (2 cm H 2 O + 8% of reading)Peak Inspiratory Pressure(PIP)Setting: 10 to 80 cm H2O (± 2 cm H2O ± 8% of setting)Setting Resolution: 1 cm H2 OMeasurement: 0 to 99 cm H 2 O ± (2 cm H 2 O + 8% of reading)