906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide 4-13SpO2 -- Using the Pulse OximeterThe primary use of the device is as a ventilator -- the pulse oximeter operates only when thedevice is providing ventilation.The following conditions can affect the pulse oximeter reading:• The sensor is too tight.• There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp,or sunlight.• A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensorattached.• The patient has hypotension, severe vascoconstriction, severe anemia, or hypothermia.• There is an arterial occlusion proximal to the sensor.• The patient is in cardiac arrest or is in shock.The SpO2 display is active only when the pulse oximeter is connected. The pulse oximeter is instandby (and displays stby in the parameter window) when• No SpO 2 sensor is connected• The sensor is off the patient during start up• You place the pulse oximeter in standbyNote: You can place the pulse oximeter in standby only when the probe is disconnected fromthe patient. A valid signal automatically brings the pulse oximeter out of standby.