906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide A-3Pulse OximeterRange Saturation (% SpO2) 1%-100%Pulse Rate (bpm) 25-240Perfusion 0.02%-20%AccuracySaturation (% SpO2)-During No Motion ConditionsAdults, Pediatrics 70%-100% 2 digits0%-69% unspecifiedNeonates 70%-100% 3 digits0%-69% unspecifiedSaturation (% SpO2)-During Motion ConditionsAdults, Pediatrics 70%-100% 3 digits0%-69% unspecifiedNeonates 70%-100% 3 digits0%-69% unspecifiedPulse Rate (bpm)-During No Motion ConditionsAdults, Pediatric, Neonates 25 to 240 3 digitsPulse Rate (bpm)-During Motion ConditionsAdults, Pediatric, Neonates 25 to 240 5 digitsResolution Saturation (% SpO2) 1%Pulse Rate (bpm) 1Low Perfusion Performance>0.02% Pulse Amplitude Saturation (% SpO2 ) 2 digitsand% Transmission >5% Pulse Rate 3 digitsInterfering SubstancesCarboxyhemoglobin may erroneously increase readings. The level of increase is approximately equalto the amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that changeusual arterial pigmentation may cause erroneous readings.