TECAN 2. General2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2-92.3.4 Tuneable Wavelength Instrument (Gradient FilterCarriage)PARAMETERS CHARACTERISTICSMeasurement time:Dual wavelengthSingle wavelength16 seconds8 secondsWavelength range:Gradient filterStandard filter400 - 700 nm340 - 399 nm & 700 - 750 nmMeasurement range:340 - 399 nm400 - 750 nm0 - 3.000 OD0 - 4.000 ODResolution: 0.001 ODAccuracy:(492 nm) 0.000 - 2.000 OD Better than +/- (1.5 % + 0.010 OD)Precision:(492 nm) 0.000 - 2.500 OD Better than +/- (1.0 % + 0.005 OD)Linearity:(492 nm) 0.000 - 2.500 OD Better than +/- 2 %Wavelength selection:Gradient filter Special gradient filter any wavelengthbetween 400 and 700 nm in 1 nm steps.The instrument can use up to eightdifferent filter carriagesFilter wavelength accuracy Central wavelength +/- 2 nmFilter bandwidthAt 50% transmission:450 nm550 nm650 nm10 - 16 nm10 - 14 nm10 - 18 nmLight source: Halogen lamp 20 WAll connected devices must be approved and listed as per EN 60950, UL 1950or CSA C22.2 No. 950 for Data Processing Devices.Computer interface:Serial RS 232 CPrinter300 - 38,400 baudOnly with the Touchscreen option.2.3.5 6 Filter Instrument (6 Filter Carriage)See 2.3.2 Standard Instrument (Standard Filter Carriage).