2. General TECAN2-12 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03Barcode labels must fulfill the following norms:• ISO/IEC 15416 Automatic identification and data capture techniques• Bar code print quality test specifications• Linear symbols (e.g. EN 1635).• ANSI X3.182-1990 (R1995): Guideline for Bar Code Print QualityNoteThe barcode must have the quality Class A, B or C / ANSI/CEN/ISOstandard. Dirty, folded, wet or damaged barcode labels must not beused. The adhesive labels must be flat and not peeling off at the edges.We recommend to guarantee the quality of the barcode labels by means of localSOP.The barcode has to be accurately aligned horizontally and positioned on the righthand side of the plate, 7 mm from the front-right edge. Position the barcode asnear as possible to the bottom.7655A1PARAMETERS CHARACTERISTICSLength max 48 mm (3.5 inch)Bars height min 6 mm (0.24 inch)Quiet zone min 5 mm (0.2 inch)Resolution min 0.15 mm (5 mil) module widthLength max 29 digits2.3.10 USB StickPlease NoteUse always the currently Tecan released and Tecan approved USBsticks! Otherwise data leakage may be possible.PARAMETERS CHARACTERISTICSStorage >= 32 MBData retention min. 2 yearsErase cycles 100,000 timesDo not expose the USB stick to extreme temperatures, fire, water, moisture, dust,smoke, electrical power and mechanical forces.