698.7.2. S-BPM OperationsNoteSet the built-in clock and initial pressurization value beforemeasurement because they are used for S-BPM. Refer to"8.2.2. The Clock and the Monitor Function ofMeasurement" and "8.4. S-BPM Programs".Remove batteries when the patient removes the recorder andcuff when ANBP or ASBP is used (even during the waitingmode). If batteries are left in the battery compartment, the cuffmay break when the recorder inflates the cuff at the next "1scope". If the patient resumes measurement, insert batteriesand press the / switch.Manual blood pressure measurement can be performed duringS-BPM waiting mode.The measurement result of the manual blood pressuremeasurement can be stored in the memory.When S-BPM is stopped, the error code E07 is displayed onthe OLED display and stored in the memory.To start S-BPMStep 1. Set the AUTO switch to "OFF".Step 2. Operations are as follows:S-BPM Programs OperationsOffice blood pressure OBPAutomated office blood pressure AOBPHome blood pressure HBPPress the / switch tostart the preset programduring waiting mode.Automated night blood pressure ANBPAutomated self blood pressure ASBPPreset program entersstandby until the "start time"or "start time of alarm".