PROPRIETARY INFORMATIONCONFIDENTIAL MATERIALSoftware Not for Customer Distribution180 AEROSET ® System Troubleshooting Guide94816-107—November 2004Rounding of NumbersThe AEROSET System allows only 7 digits, one of which maybe a decimal. When the last digit is 5 or greater, the numberrounds up. If the last digit is less than 5, the number roundsdown. For example, 1.15 rounds to 1.2 and 1.14 rounds to 1.1.Sample Sequence/Smart SamplingSmart Sampling Selected Smart Sampling DeselectedSample SequenceDefinedOnly acknowledges ICT™ assaysdefined in Sample Sequence.Aspirates photometric assays usingSmart Sampling scheduling. Iron* isalways aspirated last.Runs assays as defined in SampleSequence, except Iron. Then runsassays in the order of buttons on theASSAYS screen (left to right). Iron* islast, independent of buttonplacement.Sample SequenceNot DefinedAspirates using Smart Samplingscheduling. ICT assays are run laterin the run. Iron* is always aspiratedlast.Aspirates in the order of buttons onthe ASSAYS screen (left to right).Iron* is last, independent of buttonplacement.* Iron special handling is due to 2-cycle sample aspiration requirements.• Sample Sequence—will run first defined A-Line assay with first defined B-Line assay, etc.• Does run a B-Line assay in cuvette pair where ICT assays are run as A-Line.