PROPRIETARY INFORMATIONCONFIDENTIAL MATERIALNot for Customer DistributionAEROSET® System Troubleshooting Guide 2194816-107—November 2004General InformationAssay Configuration/InstallationRefer to the most current version of Technical Bulletin1018-2003 when installing new assays.Drugs of Abuse (DOA) AssaysThe Abbott Clinical Chemistry Drugs of Abuse (DOA) menuincludes Amphetamine/Methamphetamine, Barbiturates,Benzodiazepines, Cannabinoids, Cocaine, Ethanol,Methadone, Opiates, Phencyclidine, and Propoxyphene.Cutoff Calibrator ValuesCutoff calibrator levels used for each of the qualitative DOAassays are listed in the following table.* Cutoff concentrations mandated by SAMHSA (SubstanceAbuse and Mental Health Services Administration) for FederalWorkplace drug testing.The Ethanol assay is a quantitative assay which is calibratedusing two levels of calibrator, 0 mg/dL and 100 mg/dL.Assay LevelAmphetamine/Methamphetamine 1,000 ng/mL*Barbiturates 200 ng/mLBenzodiazepines 200 ng/mLCocaine 300 ng/mL*Cannabinoids (THC50) 50 ng/mL*Methadone 300 ng/mLOpiates 300 300 ng/mLOpiates 2,000 2,000 ng/mL*Phencyclidine (PCP) 25 ng/mL*Propoxyphene 300 ng/mL