PROPRIETARY INFORMATIONCONFIDENTIAL MATERIALNot for Customer Distribution Assays/ReagentsAEROSET® System Troubleshooting Guide 9394816-107—November 2004Procedure1. Prepare one blank level for each assay to be tested.2. Select <Database> in the Information Access Area of theMain Display. The DATABASE screen displays.3. Select <Order> in the right column of the screen. TheOrder Samples screen displays. If this is not the firstsample ordered, select <New>.4. Enter the following information:• Name or SID• C/P• Replicates (5)5. Select the assay(s) to be evaluated.6. Select <Copy> in the right column of the Order Samplesscreen. The Copy Order dialog window displays.7. Enter 1 in the How Many Copies? field to request one copy(N = 10, 2 samples × 5 replicates) then select <OK>. TheOrder Samples screen displays.8. Load the sample(s) in the designated C/P.9. Select <OK> in the right column of the screen.10. Select <RUN> in the Action Area of the Main Display. TheRUN OPTIONS screen displays.11. Select the Patient option ().12. Select <Start> in the right column of the RUN OPTIONSscreen to initiate the run.