Calibrations6H4866-02 107Section 4: CalibrationsNominal CalibrationThe Nominal Calibration procedure is identical for Kodak QC software versions 2.1.2, 2.4.2 and 2.5. The GUI isdifferent from one version to another. In this section we show the GUI of software version 2.5. Even though thescreenshots for the other software versions look different, they contain exactly the same information and instructions.Origin CalibrationImportantThe machine is calibrated at the factory before it is shipped. Sometimes it is necessary to re calibrate after a partsreplacement procedure.[1] Enter the Kodak QC configuration screen.[2] Click on the Calibration tab.Set Origin on Calibration Tab[3] Select the Set Origin option and then click Start.NoteA message with the exposure setting for the origin calibration will appear.Exposure Settings