Label Gone TimeoutProduct Reference Guide 35Label Gone TimeoutThis feature sets the time after the last label segment is seen before the reader prepares for a newlabel. The timeout can be set within a range of 10 milliseconds to 2,550 milliseconds (2.55seconds) in 10ms increments. Label Gone Timeout does not apply to scan modes that require atrigger pull for each label that is readFollow these instructions to set this feature:1. Determine the desired setting in milliseconds.2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with lead-ing zeroes to yield three digits. For example: 0 = 000, 5 = 005, 20 = 020, etc.3. Scan the ENTER/EXIT PROGRAMMING MODE barcode to enter ProgrammingMode.4. Scan the barcode: SELECT LABEL GONE TIMEOUT SETTING.5. Scan the appropriate three alpha-numeric characters from the keypad in Appendix E, Key-pad representing the duration which was determined in the steps above. You will hear atwo-beep indication after the last character.NOTE6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit ProgrammingMode.This completes the procedure. See Table 1 for some examples of how to set this feature.Table 1. Timeout Setting ExamplesSTEP ACTION EXAMPLES1 Desired Setting 50ms 150ms 1800ms (1.8sec.) 2550ms (2.55 sec.)2 Divide by 10 (and pad withleading zeroes) 005 015 180 2553 Scan ENTER/EXIT PROGRAMMING MODE4 Scan SELECT LABEL GONE TIMEOUT SETTING5 Scan Three CharactersFrom Appendix E, Keypad ‘0’, ‘0’ and ‘5’ ‘0’, ‘1’ and ‘5’ ‘1’, ‘8’ and ‘0’ ‘‘2’, ‘5’ and ‘5’6 Scan ENTER/EXIT PROGRAMMING MODEIf you make a mistake before the last character, scan the CANCELbarcode to abort and not save the entry string. You can then startagain at the beginning.