B40/B20 Patient Monitor7-14OCRGThe monitor support 8 minutes OCRG (oxycardiorespirogram) function in the NEONATAL mode.The OCRG subsystem provides services to view and review specific high resolution trends, highresolution beatto- beat HR, high resolution beat-to-beat SpO2 and compressed respirationwaveform - simultaneously in the same view. Two types of views are provided: OCRG Snapshotview and OCRG Realtime view.View OCRG snapshotThe OCRG snapshot is created when OCRG event is triggered. Refer to "Setup OCRG" for OCRGevent setup. The monitor can store up to 70 OCRG snapshots. OCRG snapshot includes thetrend data 6 minutes before and 2 minutes after the OCRG event. If multiple OCRG events aretriggered within 2 minutes, they will be detected as one OCRG event. During this time, only thefirst OCRG event will trigger OCRG Snapshot creation.To display the OCRG snapshot record.1. In NEONATAL mode, Press the Pt.Data & Trends key.2. Select OCRG Snapshots.You can select Scroll Snapshots and scroll left and right with the Trim Knob to see OCRGsnapshots earlier. The trigger time and condition also showed on the left menu area.You can select Remove Menu to hide the left side menu. Press the Trim Knob to show themenu.View realtime OCRGTo display the realtime OCRG.1. In NEONATAL mode, Press the Pt.Data & Trends key.2. Select OCRG Realtime.You can select Remove Menu to hide the left side menu. Press the Trim Knob to show the menu.Setup OCRGThe OCRG event is that any one of HR, SpO2, Apnea is out of OCRG triggering limits range. Youcan setup these limits.Adjust OCRG alarm limits1. In NEONATAL mode, Press the Pt.Data & Trends key.2. Select OCRG Setup.3. Select Adjust Limits.Adjust short apnea time1. In NEONATAL mode, Press the Pt.Data & Trends key.2. Select OCRG Setup.3. Select Short Apnea.NOTE: The default short apnea time is 20 seconds.