Alarms5-11Audio on1. Press the Alarms Setup key.2. Select Audio ON/OFF menu.3. Select Activate Alarms to turn on audible alarms.Audio paused or Alarm reset: silencing audible alarmstemporarilyNOTES:• When monitor’s remote control is set to active and the related feature has been enabledin the Central configuration. The bedside alarms can be audio paused by iCentral, themessage 'Alarms silence from Central' is displayed on the bedside monitor display. Thebedside alarms can be reset by CIC, the message 'Alarms Reset from Central' is displayedon the bedside monitor display.• If the monitor is connected to the network and the network connection is lost, the audiopaused alarms are reactivated and the volume level is automatically set to 7.* Alarms are listed as follow:− Physiology alarms:Action Result IndicatorPress audio pausekey once• Start a 2 minute audio pause periodfor all alarms except the criticalbreakthrough alarms (for example, VTach). See the “Default configurationworksheet” for details aboutbreakthrough alarm• Cease all latched alarms (includingmessage and light).• Alarm light: Yes• Alarm message: Yes and has the audiopause countdown icon• Alarm audio: No, except breakthroughalarmsPress audio pausekey second timeduring audio pausedperiod• Cease the audio pause state.• Remove alarms listed below*.• Alarm light: Yes• Alarm message: Yes• Alarm audio: YesPress alarm resetkey• Pause all active audio alarms for 2minutes.• Does not silence any new alarms.• Cease all latched alarms (includingmessage and light).• Cease the audio pause state.• Alarm light: Yes• Alarm message: Yes, with audio pausesymbol in the message block• Alarm audio: No