B40/B20 Patient Monitor5-8Adjusting alarm limitsAdjusting limits1. Press the Alarms Setup key.2. Select Adjust Limits.3. Turn the Trim knob to selest the specific alarm to be configured.4. Choose the highlight measurement. If the desired measurement is not displayed in thewindow, select Next Page.5. Push the Trim Knob to open an adjustment window.6. Turn the Trim Knob to change the limits and accept them by pushing it. Move betweenselections by turning the Trim Knob.7. To return to the Alarms Setup menu to select more measurements, push the Trim Knobuntil the cursor is in the adjustment menu, then select Previous Menu or Alarms Setup.8. Press the Normal Screen key to return to normal monitoring view.You can enter the alarm limit adjustment window also through the measurement menus.When you are in an XX parameter menu, select XX Alarm or enter XX Setup and then selectXX Alarm.NOTES:• In NIBP measurement, the alarm limits change automatically according to the cuff hosetype used.• If the monitor is connected to the network, the alarm limits can also be adjusted using theCentral if this feature has been enabled in its configuration. If alarm limits are adjustedusing the Central, the message 'Alarm settings changed from Central' is displayed in thebedside monitor.Choosing automatic limitsTo activate automatic patient-specific alarm limits enabling close patient control, select AutoLimits in the Alarms Setup menu. Limits are then calculated from the displayed patientreading at the point of time when auto limits are selected.Returning to default limitsSelect Default Limits to set the alarms to the default alarm limits. The factory default alarmlimits are listed in the "Default Configuration Worksheet."