B40/B20 Patient Monitor9-6ECG cables• Wipe the cables with mild detergent solution.• Avoid excessive use of liquids.• Disinfect when necessary.• Allow the product to dry completely after cleaning.Pulse oximetry sensorsThe GE Healthcare pulse oximetry sensors are latex-free. Take possible patient allergies intoaccount also when selecting the cleaning agent.1. Detach the sensor from the patient and the monitor.2. Wipe the sensor with mild detergent solution. Allow it to dry completely before use.Temperature probes• Clean with mild detergent solution and rinse with water.• Disinfect or sterilize when necessary.NIBP cuff hose• The cables and hose can be wiped with mild detergent solution.• Disinfect when necessary.NIBP cuffThe NIBP cuffs listed in the “Supplies and Accessories” catalog are latex-free. Take possiblepatient allergies into account also when selecting the cleaning agent.• Clean only when necessary. Wash the cuff in mild detergent solution. Do NOT use alcohol.Invasive blood pressure cables• Wipe the cables with sterile alcohol-based detergent. After cleaning rinse surfaces bywiping them with a cloth damped with sterile water. Dry with a dry cloth.Airway adapter• Replace the single use adapter after each patient.Sampling line• Do not reuse the sampling line. Reusing a cleaned sampling line may affect measurementresults.NOTE: Do not wash or disinfect calibration gas sampling lines.Water trapWARNING Do not disinfect or open the water trap cartridge. Do not touch the watertrap membrane. The hydrophobic membrane is damaged if any cleaning isattempted other than rinsing with water.The water trap is based on a hydrophobic membrane, which prevents water and secretionsfrom entering the measuring chamber. Condensed water and saliva are collected into awashable container.• Empty the water trap container as soon as it is more than half full. With a sample gastemperature of 37°C, a room temperature of 23°C, and sample gas relative humidity of100 %RH, the water trap should be emptied every 24 hours (applies when the sample gas