Media-Mate Plus Operation Manual 25-710-800-OSection Five – InstallationLocationEnvironmentThe media tank temperature must exceed ambient room temperature by at least 5°C formaximum performance of the temperature control system. Therefore, for best results, theambient temperature should not be allowed to exceed 32°C (90°F) for 37°C tests, or 27°C(81°F) for 32°C tests.Easy access to a water source and waste disposal for filling and emptying of the waterbath isconvenient, although not necessary.SpaceThe Media-Mate Plus System is designed to be mobile up to 6.25m (20 ft). The laboratory floorshould be level and flat to prevent the Instrument from moving when left unattended.ElectricalThe Media-Mate Plus System requires a single, grounded electrical outlet, within 5.63 m (18 ft)of the location of the Instrument.The Instrument is available in two configurations, 115VAC and 230VAC. The voltages are setat the factory and cannot be changed (see model number label on control interface panel nextto fuses to determine voltage). The electrical requirements are summarized in Section 3 of thisoperation manual.UnpackingThe Media-Mate Plus is shipped fully assembled in one box. The shipping box is specificallydesigned to provide maximum shipping protection and to facilitate unpacking. The shippingbox consists of a lower cardboard tray secured to a pallet with a large box placed over theInstrument. The upper box is strapped to the lower tray and pallet.NOTE: Shortages or damages must be reported immediately to the freight carrier and to HansonResearch. Notify Hanson Research by telephone (818) 882-7266 or FAX (818) 882-9470.To Unpack:1. Move package close to final destination with pallet-moving equipment.2. Cut and remove straps.3. Remove upper box taking care not to damage box.4. Remove Instrument from box.5. Reassemble empty box and store for possible future use.Section 5 Installation 9