Media-Mate Plus Operation Manual 25-710-800-OInstallationCAUTIONThe Instrument should not be plugged in until installed as per the instructions below.1. Clean the inside of the Tank as necessary to remove any packing debris and fill withapproximately 40L of DI water (see Section 9 – Cleaning the Tank and Fluid Lines).2. Install the Heater Assembly into the Tank as shown (see Figure 5-2). Place Tank Coveronto Tank (see Figure 5.5).CAUTIONThe filled Tank is heavy and solution may spill. This installation requires two people. Theperson holding and pushing the Tank onto the Cart legs should be capable of supportingthe weight of the Tank.3. One person aligns the Media-Mate Plus cart legs with steps on Tank from the back. Tipthe Tank just enough so that the Cart legs can be pushed under the Tank. Lift the oppositeside of the Tank and push as you slide the Cart legs under.HINT: Keep the front wheels of the Cart straight.4. Remove the Fill Hose Assembly from its holding position on the side of the Tank andconnect to the Fill Manifold (see Figure 5-2).5. Connect the Heater connector into the Interface Panel (see Figure 5-4).6. Connect the Dispense Tubing to the Dispense Head and position over the holes in theTank Cover (see Figure 5-5).NOTE: Configure the Dissolution Test Station Vessel Pattern (see Section 7 – Configuringthe Dispense Head).Section 5 Installation 14