Media-Mate Plus Operation Manual 25-710-800-OStart UpCAUTIONThe voltage indicated on the Media-Mate Plus System must be the same as the power source.If the incorrect voltage is indicated, do not plug in the Instrument and contact Hanson Researchimmediately.CAUTIONDo not press the Start Switch until the indicator lights are off as this may cause overfilling ofthe System.The Instrument must be plugged into a grounded power outlet.1. Pull the power cord out of the retractable cord reel and plug the other end of the powercord into a grounded three-prong power outlet.NOTE: The power cord plug may need to be replaced for some countries’ power outletconfigurations. Ensure that the replacement plug is approved and properly ratedfor voltage and amperage by following the Power Cord Color Code below:USAColor CodeWireDescriptionInternationalColor CodeBlack Live BrownWhite Neutral Light BlueGreen Ground Green with Yellow Stripe2. Turn on the Main Power Switch, located on the Pump Panel next to the power cord outlet.3. Turn on the Air Pump and Heater power switches, located on the Control Panel.4. Run a Test Cycle (see Section 7 – Test Cycle) while observing Instrument performance.5. During the Test Cycle, verify the correct sequence of events by observing the SequenceLEDs on the Control Panel and by visually observing fluid flow at the measure vessels.6. During the Test Cycle, verify Heater operation by observing the “1” LED on theTemperature Controller (“1” is energized when the Heater is on).7. Turn off the Heater and Air Pump switches.8. Turn off the Main Power Switch and remove the power cord from the grounded three-prong power outlet and return into retractable cord reel.Section 5 Installation 15