CDE/CME/CDEE-BA-e-1014 258.4.4 Selection of the adjustment weightFor the model ranges KERN CDE and CME the adjustment weight can be selectedfrom three or four preset nominal values (c.1/3; 2/3; max or c.1/4; 1/2; 3/4; max) (seealso table 1 below, default settings are highlighted in grey). In order to achieve high-quality weighing results in the sense of the measuring technology, it is recommendedto select the nominal value as high as possible.Ö With the weighing scale switched off, press ON/OFF and TARE-keysimultaneously until the first function “Unit“ appears.Ö Repeatedly press the 10x-key until “CAL“ appears.Ö Confirm by pressing the PRINT-key. Current setting appears.Ö Use the 10x-key, to select among the preset nominal values (seetab. 1)Ö Confirm your selected setting by pressing the PRINT-key.CME 100-3 CME 300-2 CME 1000-2 CME 3000-1 CME 6000-120 g 100 g 200 kg 1 kg 2 kg50 g 200 g 500 kg 2 kg 5 kg100 g 300 g 1000 kg 3 kg 6 kgCDE 35K1 CDE 35K1L CDE 60K2 CDE 60K2L CDE 150K5 CDE 150K5L5 kg 5 kg10 kg 10 kg 20 kg 20 kg 50 kg 50 kg20 kg 20 kg 40 kg 40 kg 100 kg 100 kg30 kg 30 kg 60 kg 60 kg 150 kg 150 kg8.4.5 Reset to factory settingWith this function the manual changes of the menu settings are reset to factorysetting.Ö With the weighing scale switched off, press ON/OFF and TARE-keysimultaneously until the first function “Unit“ appears.Ö Repeatedly press the 10x-key until “rSt“ appears.Ö Confirm by pressing the PRINT-key. Current setting appears.Ö Now you can use the 10x-key to select among the settings below:„rSt no“: No reset to default setting„rSt yes“: Reset to default settingÖ Confirm your selected setting by pressing the PRINT-key.