CDE/CME/CDEE-BA-e-1014 3512.4 Installation• Switch off both scales and disconnect• Connect both scales with the data cable provided for the RS232 interface12.5 Balance settingsTo enable automatic transfer of the reference weight from CME to CDE, the datatransfer mode “Pr” in the menu (see chpt.9.3.1) must be set to “CSyS“.12.6 Counting with both scalesAttention:To prevent errors during the piece count determination, both scales must be adjustedto the same gravity acceleration (see chap. 6.8).Non-compliance will result in counting errors!1. Place empty vessel on weighing platform of quantity scale KERN CDE and tareby pressing the key (TARE)2. Place reference piece number of count quantity on reference scale KERN CME(e.g. 5, 10, 20 items)3. Confirm selected reference piece number by pressing the key (5, 10, 20) on thereference scale.4. Fill vessel on quantity scale with count quantity. The piece number is showndirectly on the display. The -key may be used to switch over to piece weight.