8.10 Tolerance weighingFor tolerance weighing compare the current weight value to a lower and upper limit.These limits must have been stored beforehand.When placing for instance ready-made packaging, it will be shown whether theweight is within the allowable tolerance.Operation: Display:Ö Turn on display unit by pressing the ON/OFF key 0.00 gÖ Press F -key 1 FuniÖ Repeatedly press the a key until 3 HiLo appears 3 HiLoÖ Press TARE key Min 0000.00gÖ To enter the lower tolerance limit (e. g. 90.00 g): Use theTARE key, to select the digit to be changed, the a keyis used to increase the numerical value (active digitflashing)Min 0090.00gÖ Press the PRINT key; the lower tolerance limit will beimported.Ö To enter the upper tolerance limit (e. g. 110.00 g): Use theTARE key, to select the digit to be changed, the a keyis used to increase the numerical value (active digitflashing)Max 0100.00gÖ Press the PRINT key; the upper tolerance limit will beimported and jumps to tolerance weighing mode.Ö Example for application: Weight: 50 g on weighing plateÖ Weight: 100g on weighing plateÖ Weight: 150g on weighing plateMin 50.00gOk 100.00gMax 150.00gÖ Go back to weighing mode by pressing the F key twice. 0.00 gThe value entered for the upper tolerance limit must be greater than the lowertolerance value. Error message –Lo- appears on the display and the display of theunit returns to weighing mode.KLB-TM-BA-e-081134