12 Instant helpIn case of an error in the program process, briefly turn off the display unit anddisconnect from power supply. The weighing process must then be restarted from thebeginning.Help:Fault Possible causeThe displayed weight doesnot glow. • The display unit is not switched on.• The mains supply connection has been interrupted(mains cable not plugged in/faulty).• Power supply interrupted.• (Rechargeable) batteries are inserted incorrectly orempty• No (rechargeable) batteries inserted.The displayed weight ispermanently changing • Draught/air movement• Table/floor vibrations• Weighing plate has contact with other objects.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)The weighing result isobviously incorrect • The display of the balance is not at zero• Adjustment is no longer correct.• Great fluctuations in temperature.• Warm-up time was ignored.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)Should other error messages occur, switch display unit off and then on again. If theerror message remains inform manufacturer.KLB-TM-BA-e-0811 47