9.8 Communication protocol / remote control commandsTable 1: Instruction list for RS232 interfaceInstruction: Meaning of Instruction:Z Set weight display at zeroT TaringS Send stable weighing value in standard weighing unitSI Send weighing value immediately in standard weighing unitSU Send stable weighing value in current weighing valueSUI Send weighing value immediately in current weighing unitC1 Turn on continuous transmission in standard weighing unitC0 Turn off continuous transmission in standard weighing unitCU1 Turn on continuous transmission in current weighing unitCO1 Turn off continuous transmission in current weighing unitPC Send all implemented instructionsComplete each instruction with CR LF.9.8.1 Response messages from balanceResponse message of balance after instruction was sent:XX_ Instruction:XX_A CR LF Instruction accepted; will be executedXX_D CR LF Instruction complete (appears after XX_A only)XX_ I CR LF Instruction received; impossible to carry outXX_^ CR LF Instruction received but time overflow error occurredXX_v CR LF Instruction received, but insufficient loadXX_E CR LF Error during execution, timeout for stable weighing valueexceededKLB-TM-BA-e-0811 43