Refer to the RP793xA, RP794xA and RP795xA, RP796xA characteristics tables for the model-specificresistance programming ranges.Front Panel Menu Reference SCPI CommandSelect Output\Advanced\Resistance.Specify an Output Resistance value. Then check theEnable box. Then press Select.To select a resistance of 0.5 ohms:VOLT:RES 0.5To enable output resistance:VOLT:RES:STAT ONWhen units are paralleledd, the programmable output resistance is reduced. Theprogrammable output resistance for a single unit must be divided by the totalnumber of paralleled units.Set the Output Bandwidth - RP795xA, RP796xAThis information applies to models RP795xA and RP796xA only.Voltage compensation modes let you optimize output response time with capacitive loads.High1 compensation setting provides maximum up-programming speed as well as the fastesttransient response settling time. This mode is optimized for use with resistive loads; howevercapacitive loads up to the limits outlined in the table below can be used when using load leads shorterthan 3 meters (10 feet). Exceeding these limits can lead to voltage programming overshoots andtransient response instability.Low compensation setting is optimized for use with large capacitive loads up to the limits shown in thetable below, along with load leads longer than 3 meters (10 feet). In this mode, the up- and down-programming speed and voltage control-loop bandwidth are limited to prevent voltage programmingovershoots and improve transient response stability. Low mode provides the best stability andovershoot minimization over all load configurations. specifies the low-pass corner frequency of a filter applied to the programming signal.Refer to the frequency description in the next sectionConnecting very low ESR capacitors larger than the High mode limit with load leadsshorter than 3 m (10 ft.) is not recommended in either bandwidth range. This loadconfiguration can lead to voltage programming overshoots.Setting 5 kW Models 10 kW Models 950 V ModelsHigh1 0 to 80 μF 0 to 80 μF 0 to 40 μFLow 0 to 100,000 μF 0 to 100,000 μF 0 to 50,000 μFKeysight RP7900 Series Operating and Service Guide 1054 Using the Regenerative Power System