C-3C.3 Chiller Setup and OperationC.3.1 SetupYour Labcyte service representative will set up the chiller during Echosystem installation. If you need to set up the chiller again (for example,after relocation of the Echo system), use the following procedure.The Echo system should already be set up, with tubing assembly attachedto the fluid panel (see Echo liquid handler user manual).1. Place the chiller on the benchtop near the Echo unit, or on the floorunder the Echo unit.2. Connect the Echo tubing assembly to the fluid connections on the chillerlabeled Coolant Return for liquid returning to the chiller and CoolantSupply for fluid exiting the chiller.Figure C.2 Tubing assembly connections3. Connect the power supply cable: circular 2-pin connector to the 12 V DC (8 A) on the chiller. three-prong electrical connector to the electrical outlet.