14.2.2 Module-specific critical errorsLog entry Parameter DescriptionCode Plain text Code Plain text2002 MODULE ERROR(002)0 --- There was a memory overflow in one of the twoinfinite counters.2003 MODULE ERROR(003)0 --- Internal errorContact the manufacturer if the device logincludes this error several times (see "Branchoffices" on page 141).2004 MODULE ERROR(004)0 --- The chargeable "Unlocking of integrated RS232interface (NLM0V3RS232)" feature required touse the optional GNSS receiver has not beenreleased.2005 MODULE ERROR(005)0 --- Timeout when accessing the optional GNSSreceiver14.2.3 Module-specific information regarding the current operating stateLog entry Parameter DescriptionCode Plain text Code Plain text2400 MODULE INFO(2400)1 --- Additional connection establishment initiatedindependently of the transmission cycle2402 MODULE INFO(2402)## --- Operating mode changedThe parameter specifies the new operatingmode (1 = normal, 2 = Aloha, 3 = transportmode, 4 = production mode).2403 MODULE INFO(2403)0 --- The power supply was switched back to supplyvoltage.1 --- The power supply was switched to therechargeable buffer battery.14.3 Evaluating the device log14.3.1 Evaluating the device log on the Device to Web serverThe last 300 log entries on the Device to Web server can be called up via the button shown below that islocated in the measurement device list. As the log entries are sent to the server in the transmission cycle in thesame way as the measurement data, only the log entries up to the last server connection are available.The manual for the server ("Device to Web Server Manual" - upon request) includes a detailed description ofthe evaluation of the device log on the Device to Web server.128 of 141 NivuLog Easy V3 - Rev. 03 as of 24.10.2018