114 InstallationInstall the ANUJ5024 Aicure under the conditions below.4.1 Installation conditions1) Ambient temp.: Main unit: 0 to +35°CHead: +5°C to +35°C2) Relative humidity: No greater than 85%RH, with no condensation3) Make sure the rubber legs are set horizontally parallel.4) In order to avoid damage due to overheating, do not use if the areaaround the system is blocked.5) The dimensions of the main unit are 80 mm (L) x 130 mm (H) x 145mm (D), but you should take into account the state of the unit withan LED head attached.The minimum allowable bending radius for the LED head cord is 33mm. Using the cord with a curve radius of less than 33 mm coulddamage it.Do not place items on top of the main unit, or block its air vents.Excessive heat could damage the unit due to overheating.WarningWhen using more than one controller, donot bundle the supplied AC adapterstogether. The AC adapters may overheatand damage the products.Secure enough spaceabove and on the sides ofthe controller.WallApprox. 50 mmApprox. 20 mmWallWarningWarning*The power cable supplied with thisproduct is a 100VAC cable for usein Japan only. When using thisproduct outside Japan, use apower cable that has beencertificated by applicablestandards and has a plug shapethat fits the receptable used in eachcountry.