Chapter 8: Installing, Removing, and Replacing ComponentsRemoving and Replacing the System Control BoardQuantum Scalar i40 and Scalar i80 User’s Guide 2295 Check the library’s date, time, and time zone settings andreconfigure if necessary. From the Web client, select Setup > Date& Time.6 If you changed the IP address from what it was originally, updatehost and other applications that access the library.7 Save the library configuration. From the Web client, select Tools >Save/Restore Configuration. Select Save Configuration and clickApply. Save the configuration to a known location.8 It is recommended that you run the Installation and Verification Test(IVT). From the operator panel, select Tools > IVT and follow theinstructions. The IVT takes about 30 minutes. If any problems aredetected, the library issues a diagnostic ticket.9 Package the old SCB in the packaging that shipped with the newSCB and send the old SCB back to Quantum.Manually Reconfiguringthe LibraryFollow the instructions in this section if you do not have a saved libraryconfiguration.1 If you had licenses installed on your library, have the license keyshandy so you can install them on the library. If you no longer havethem, you can obtain them at http://www.quantum.com/licensekeys, or from Quantum Support.2 When the firmware update is complete, the Setup Wizard appearson the operator panel screen. The Setup Wizard helps you configurethe date, time, IP address, IPv6 enable/disable, licenses, partitions, I/E slots, and cleaning slots.3 Complete the Setup Wizard on the operator panel. If you make amistake or do not complete all the screens, you can change theconfiguration later using the operator panel or Web client Setupmenus.Note: You cannot update the library name or time zone on theoperator panel. If the displayed library name or time zoneis incorrect, change it later using the Web client (see Step 8on page 230).4 Take note of the library’s IP address (on the operator panel, selectReports > Network Settings).