Start-UpAxiostar plus Attaching epi-fluorescence illuminator Carl ZeissB 40-815 e 12/01 2-252.11.3 Replacing burner in fluorescentlamp• Turn off power supply for HBO 50. RemoveHBO 50 plug from the socket on the powersupply.Danger of getting burned! Allow lamp housingto cool off at least 15 minutes.• Remove HBO 50 lamp from the epi-fluorescence illuminator and place it on awork surface.• Loosen clamping screw (2-23/1) with Allenkey (SW 3) and lift up lamp housing.• Press spring handle down (2-23/4) and liftup heat sink (2-23/2) with the burner(2-23/3) out of the lamp frame (2-23/5). Setit down so that clamping screw on the heatsink is accessible.The cable on the heat sink cannot beremoved.• Loosen clamping screw on the heat sinkwith the Allen key and remove used burner.Dispose of used burner according to applicableordinances. Pay attention to manufacturer‘ssuggestions.1 Clamping screw2 Heat sink3 Brenner4 Spring handle5 Lamp frameFig. 2-23 Replacing burnerArtisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com