Start-UpCarl Zeiss Attaching epi-fluorescence illuminator Axiostar plus2-30 B 40-815 e 12/011 Fastening ring2 Locking filter3 P&C reflector module4 Orientation groove5 Excitation filter6 Mounting sheetFig. 2-27 Changing filter set in P&C reflector module2.11.6 Changing filter set in P&Creflector moduleYou can combine and attach the filter sets forthe P&C reflector module as needed. Filter setsor completely equipped P&C reflector modulescan be ordered from Carl Zeiss.• Remove P&C reflector module (2-27/3) fromthe reflector revolver and place in a secureposition.• Unscrew fastening ring (2-27/1) withmounting sheet (2-27/6) from the tool set.• Turn over reflector module so that the filter(2-27/2 or 5) falls onto a soft surface.• Insert the locking filter (emission filter)(2-27/2) and excitation filter (2-27/5) andsecure with the fastening (2-27/1).The emission filter and excitation filter will belabeled with an inscription or arrow on theperimeter. The arrow indicates the installationdirection for the filter into the reflector moduleand should point in (see arrow in Fig. 2-27).To minimize image shift during multiplefluorescence procedures, the emission filter canhave an additional marking showing theposition of the wedge angle.The marking should be aligned to theorientation groove (2-27/4) when inserting anemission filter in the reflector module. Thisguarantees that the wedge angle of theemission filter has the same defined position inthe reflector modules used. This compensatesfor or minimizes the already slight image shiftwhen Zeiss filter sets are used betweenmodules.Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com