214.3.4.2 Set DateHighlight the set date option and press the Enter key to confirm. This will allow the user to enter thecurrent date in the format DD.MM.YY using the numerical keypad and the decimal point keys. Pressthe Enter key to confirm. Once complete the instrument will return to the Clock Setup menu.4.3.4.3 Display Time / DateHighlight the display time or display date option and press the Enter key to confirm. This will set thetop right hand corner to show the selected option.4.3.5 Refresh Dark CurrentHighlight the refresh dark current option and press the Enter key to confirm. This will re-measure andstore the detectors’ dark current values.4.3.6 Connect to PCConnect the interface cable to the USB type B port on the rear of the instrument and connect to aUSB port on the PC. The connecting COM port should be configured with the following settings:38400 baud 8 data bitsNo parity 1 stop bitHighlight the Connect to PC option and press the Enter key to confirm.4.3.7 Beeper On/OffHighlight the beeper on/off option and press the Enter key to confirm. An audible sound will be heardeach time an instrument button is pressed when this option is enabled.4.3.8 Language SelectionNot currently available.4.3.9 Refresh System BaselineEnsure the sample chamber is empty, highlight the refresh system baseline option and press theEnter key to confirm. This will re-measure and store the detectors’ signals over the full wavelengthrange of the instrument.